MEAN WELL
MPS-30-48
Single-output medical type power supply unit (PSU) - Meanwell (MPS-30) - output voltage 48Vdc - Output current 0.6A - Rated power 30W - Input voltage 88-264Vac (100Vac / 110Vac / 200Vac / 208Vac / 220Vac / 230Vac / 240Vac nom.) (Single phase) - Rated for -10…+60°C ambient - 1-output
Please add this item to cart to request a quote or contact us at [email protected] for product availability.
Brand
SKU
MPS-30-48
Please add this item to cart to request a quote.
Please add this item to cart to request a quote or contact us at [email protected] for product availability.
MPS-30-48Meanwell - MPS-30
Description
Single-output medical type power supply unit (PSU) - Meanwell (MPS-30) - output voltage 48Vdc - Output current 0.6A - Rated power 30W - Input voltage 88-264Vac (100Vac / 110Vac / 200Vac / 208Vac / 220Vac / 230Vac / 240Vac nom.) (Single phase) - Rated for -10…+60°C ambient - 1-output
Additional Information
MEAN WELL OPEN FRAME; HEALTHCARE APPROVED
Specifications
Primary brand
Meanwell
Main function
Power supply unit
Product series / family name
Open Frame
Sub-range name
MPS-30
Functions
Single-output medical type power supply unit (PSU)
1-output
1-output
Supply voltage (AC)
88-264Vac (100Vac / 110Vac / 200Vac / 208Vac / 220Vac / 230Vac / 240Vac nom.)
Supply voltage (DC)
120-370Vdc (120Vdc / 127Vdc / 200Vdc / 220Vdc / 250Vdc nom.)
Type of network
Single phase
Rated current
0.6A
Rated power
30W
Output voltage (DC)
48Vdc
Protection functions
Short-circuit
Over load (Rated output power)
Over voltage
Over temperature
Over load (Rated output power)
Over voltage
Over temperature
Protection settings
Over load (>105%)
Over voltage (110-135%)
Over voltage (110-135%)
Efficiency
0.78
Leakage current
0-0.2 mA
Ambient air temperature for operation
-10…+60 °C
Application type
Medical type
Manufacturer product status
Commercialized
Compliant with standard(s)
ANSI/AAMI ES60601-1
IEC60601-1
IEC55011
IEC61000-3-2
IEC61000-3-3
IEC61000-4-2
IEC61000-4-3
IEC61000-4-4
IEC61000-4-5
IEC61000-4-6
IEC61000-4-8
IEC61000-4-11
IEC60601-1-2
2xMOOP
IEC60601-1
IEC55011
IEC61000-3-2
IEC61000-3-3
IEC61000-4-2
IEC61000-4-3
IEC61000-4-4
IEC61000-4-5
IEC61000-4-6
IEC61000-4-8
IEC61000-4-11
IEC60601-1-2
2xMOOP
Accuracy class
0.02
Ripple
240mV
Minimum temperature
-10 °C
Maximum temperature
60 °C
Related Documents
Name | Last Update | Language |
---|---|---|
Meanwell - MPS-30 - Datasheet | 02 Oct 2017 | English (US) |